Efficacy and safety of bipolar plasma vaporization of the prostate with "button-type" electrode compared with transurethral resection of prostate for benign prostatic hyperplasia - Abstract

BACKGROUND: Transurethral resection of prostate (TURP) has been considered as the standard treatment for benign prostatic hyperplasia (BPH).

However, issues that have not yet been overcome for TURP include bleeding and absorption of irrigation fluid. Thus, novel improvement of the surgery is necessary. This study aimed to evaluate the efficacy and safety of bipolar plasma vaporization of the prostate (BPVP) with "button-type" electrode against standard TURP for BPH.

METHODS: From January 2009 to January 2012, 30 patients who scheduled for surgical treatment of BPH surgical treatment were enrolled in the trial with provided consent for the study. Patients were prospectively randomized 1:1 to undergo either BPVP or TURP. Participants were blinded to the randomization scheme. All cases were assessed preoperatively and followed at 1, 3, and 6 months postoperatively by indwelling catheter time, blood loss, hospital stays, International Prostate Symptom Score (IPSS), quality of life (QOL), and Qmax.

RESULTS: BPVP was significantly superior to TURP in terms of indwelling catheter time ((4.1 ± 4.1) days vs. (6.8 ± 6.8) days, P = 0.000), blood loss ((64.7 ± 103.8) ml vs. (254.7 ± 325.4) ml, P = 0.040), hospital stay ((8.7 ± 1.0) days vs. (11.7 ± 1.5) days, P = 0.000), IPSS ((4.2 ± 8.0) vs. (9.3 ± 3.7), P = 0.049), QOL ((1.5 ± 0.8) vs. (2.6 ± 1.4), P = 0.027), Qmax ((16.3 ± 5.7) ml/s vs. (12.5 ± 3.1) ml/s, P = 0.038), hemoglobin ((130.7 ± 9.4) g/L vs. ((122.1 ± 11.9) g/L, P = 0.047), Na+ level ((138.6 ± 2.1) mmol/L vs. ((137.2 ± 2.0) mmol/L, P = 0.046) and operation time ((39.0 ± 15.5) minutes vs. ((69.3 ± 24.8) minutes, P = 0.004). And there were no statistical differences between BPVP group and TURP group in preoperatively assessment: patient's age ((70.9 ± 7.1) years vs. (71.9 ± 6.1) years, P = 0.736), IPSS ((24.6 ± 4.7) vs. (27.3 ± 5.9), P = 0.100), QOL ((5.1 ± 0.8) vs. (5.1 ± 1.0), P = 0.940), Qmax ((4.4 ± 2.7) ml/s vs. (5.3 ± 2.6) ml/s, P = 0.314), hemoglobin ((137.4 ± 8.7) g/L vs. (139.2 ± 10.4) g/L, P = 0.623), Na+ level ((140.5 ± 1.8) mmol/L vs. (141.3 ± 1.4) mmol/L, P = 0.192) and prostate volume ((59.0 ± 17.4) ml vs. (70.1 ± 28.8) ml, P = 0.276).

CONCLUSIONS: Compared with TURP, BPVP with "button-type" electrode shows superior efficacy and safety. Therefore, BPVP with "button-type" electrode represents a valuable endoscopic treatment alternative for BPH patients.

Written by:
Zhang SY, Hu H, Zhang XP, Wang D, Xu KX, Na YQ, Huang XB, Wang XF.   Are you the author?
Department of Urology, Peking University People's Hospital, Beijing 100044, China.

Reference: Chin Med J (Engl). 2012 Nov;125(21):3811-4.


PubMed Abstract
PMID: 23106879

UroToday.com BPH Section