A feasibility study of novel ultrasonic tissue characterization for prostate-cancer diagnosis: 2D spectrum analysis of in vivo data with histology as gold standard - Abstract

Department of Radiation Oncology, Emory University School of Medicine, Atlanta, Georgia 30322, USA.

This study demonstrates the feasibility of using a novel 2D spectrum ultrasonic tissue characterization (UTC) technique for prostate-cancer diagnosis. Normalized 2D spectra are computed by performing Fourier transforms along the range (beam) and the cross-range directions of the digital radio-frequency echo data, then dividing by a reference spectrum. This 2D spectrum method provides axial and lateral information of tissue microstructures, an improvement over the current 1D spectrum analysis which only provides axial information. A pilot study was conducted on four prostate-cancer patients who underwent radical prostatectomies. Cancerous and noncancerous regions of interest, identified through histology, were compared using four 2D spectral parameters: peak value and 3 dB width of the radially integrated spectral power (RISP), slope and intercept of the angularly integrated spectral power (AISP). For noncancerous and cancerous prostatic tissues, respectively, our investigation yielded 23 +/- 1 and 26 +/- 1 dB for peak value of RISP, 7.8 +/- 0.5 degrees and 7.6 +/- 0.6 degrees for 3 dB of RISP, -2.1 +/- 0.2 and -2.7 +/- 0.4 dB/MHz for slope of AISP, and 92 +/- 5 and 112 +/- 6 dB for intercept of AISP. Preliminary results indicated that 2D spectral UTC has the potential for identifying tumor-bearing regions within the prostate gland.

Written by:
Liu T, Mansukhani MM, Benson MC, Ennis R, Yoshida E, Schiff PB, Zhang P, Zhou J, Kutcher GJ.   Are you the author?

Reference:
Med Phys. 2009 Aug;36(8):3504-11.

PubMed Abstract
PMID:19746784

UroToday.com Prostate Cancer Section