Centre for Advanced Imaging, University of Queensland, Brisbane, Australia.
Prostate cancer detection using diffusion-weighted imaging is highly affected by the accuracy of the apparent diffusion coefficient (ADC) values in an image. Echo planar imaging (EPI) is a fast sequence commonly used for diffusion imaging but has inherent magnetic susceptibility and chemical shift artefacts associated. A diffusion sequence that is less affected by these artefacts is therefore advantageous. The half-Fourier acquisition single-shot turbo spin-echo (HASTE) sequence was chosen. The diffusion sequences were compared in image quality, repeatability of the ADC value and the effect on the ADC value with varied b values. Eight volunteers underwent three scans of each sequence, on a 1.5-T Siemens system, using b values of 0, 150, 300, 450, 600, 750, 900 and 1000 s/mm2. ADC maps were created to address the reproducibility of the ADC value when using two b values compared to eight b values. The ADC value using all b values with the HASTE sequence gave the best performance in all tested categories. Both sequences gave significantly different ADC mean values for two b values compared to when using eight b values (P< .05) suggesting larger error is present when using two b values. HASTE was shown to be an improvement over EPI in terms of repeatability, signal variation within a region of interest and standard deviation over the volunteer set. The improved accuracy of the ADC value in the HASTE sequence makes it potentially a more sensitive tumor detection technique.
Written by:
Babourina-Brooks B, Cowin GJ, Wang D. Are you the author?
Reference: Magn Reson Imaging. 2011 Nov 4. Epub ahead of print.
doi: 10.1016/j.mri.2011.09.024
PubMed Abstract
PMID: 22055748
UroToday.com Prostate Cancer Section