Real-time in vivo rectal wall dosimetry using plastic scintillation detectors for patients with prostate cancer - Abstract

We designed and constructed an in vivo dosimetry system using plastic scintillation detectors (PSDs) to monitor dose to the rectal wall in patients undergoing intensity-modulated radiation therapy for prostate cancer. Five patients were enrolled in an Institutional Review Board-approved protocol for twice weekly in vivo dose monitoring with our system, resulting in a total of 142 in vivo dose measurements. PSDs were attached to the surface of endorectal balloons used for prostate immobilization to place the PSDs in contact with the rectal wall. Absorbed dose was measured in real time and the total measured dose was compared with the dose calculated by the treatment planning system on the daily computed tomographic image dataset. The mean difference between measured and calculated doses for the entire patient population was -0.4% (standard deviation 2.8%). The mean difference between daily measured and calculated doses for each patient ranged from -3.3% to 3.3% (standard deviation ranged from 5.6% to 7.1% for four patients and was 14.0% for the last, for whom optimal positioning of the detector was difficult owing to the patient's large size). Patients tolerated the detectors well and the treatment workflow was not compromised. Overall, PSDs performed well as in vivo dosimeters, providing excellent accuracy, real-time measurement and reusability.

Written by:
Wootton L, Kudchadker R, Lee A, Beddar S.   Are you the author?
Department of Radiation Physics, The University of Texas M D Anderson Cancer Center, 1515 Holcombe Blvd, Houston, TX, USA.

Reference: Phys Med Biol. 2014 Feb 7;59(3):647-60.
doi: 10.1088/0031-9155/59/3/647


PubMed Abstract
PMID: 24434775

Go "Beyond the Abstract" - Read an article written by the authors for UroToday.com

UroToday.com Prostate Cancer Section