Volumetric-modulated arc therapy (VMAT) is a widespread intensity-modulated radiation therapy (IMRT) method, however, VMAT requires adaptation of the radiation treatment planning system (RTPS) and linear accelerator (linac); these upgrades are quite expensive.
The Smart Arc of Pinnacle3 (Philips), which is the software used in VMAT calculations, can select constant dose rate (CDR) mode. This approach has a low initial cost because the linac upgrade is not required. The objective of this study was to clarify the utility of CDR mode for prostate IMRT. Pinnacle3 and Clinac 21EX linac (Varian, 10 MV X-rays) were used for planning. The plans were created for 28 patients using a fixed multi-field IMRT (f-IMRT), VMAT and CDR techniques. The dose distribution results were classified into three groups: optimal, suboptimal and reject. For the f-IMRT, VMAT and CDR results, 25, 26 and 21 patients were classified as 'optimal', respectively. Our results show a significant reduction in the achievement rate of 'optimal' for a CDR when the bladder volume is < 100 cm3. The total numbers of monitoring units (MUs) (average ± 1σ) were 469 ± 53, 357 ± 35 and 365 ± 33; the average optimization times were ∼50 min, 2 h and 2 h 40 min, and the irradiation times were ∼280 s, 60 s and 110 s, respectively. CDR can reduce the total MUs and irradiation time compared with f-IMRT, and CDR has a lower initial cost compared with VMAT. Thus, for institutions that do not currently perform VMAT, CDR is a useful option. Additionally, in the context of patient identification, bladder volume may be useful.
Written by:
Hatanaka S, Tamaki S, Endo H, Mizuno N, Nakamura N. Are you the author?
Kanagawa Cancer Center, 2-3-2, Nakao, Asahi-Ku, Yokohama City, Kanagawa, 241-8515, Japan.
Reference: J Radiat Res. 2014 Feb 11. Epub ahead of print.
doi: 10.1093/jrr/rrt232
PubMed Abstract
PMID: 24522268
UroToday.com Prostate Cancer Section